Journal article

Numerical investigations of the haemodynamic changes associated with stent malapposition in an idealised coronary artery

EKW Poon, P Barlis, S Moore, W PAN, Y Liu, Y YE, Y XUE, SJ Zhu, A Ooi

Journal of Biomechanics | Pergamon-Elsevier Science | Published : 2014

Abstract

The deployment of a coronary stent near complex lesions can sometimes lead to incomplete stent apposition (ISA), an undesirable side effect of coronary stent implantation. Three-dimensional computational fluid dynamics (CFD) calculations are performed on simplified stent models (with either square or circular cross-section struts) inside an idealised coronary artery to analyse the effect of different levels of ISA to the change in haemodynamics inside the artery. The clinical significance of ISA is reported using haemodynamic metrics like wall shear stress (WSS) and wall shear stress gradient (WSSG). A coronary stent with square cross-sectional strut shows different levels of reverse flow fo..